Documenting Process Calibrator
Specifications
91
DC Voltage Output
RANGE RESOLUTIONS
MODEL 702
% OUTPUT +% OF FULL SCALE
MODEL 701
% OUTPUT + % OF FULL SCALE
1 YEAR 2 YEAR 1 YEAR 2 YEAR
110 mV 1 µV 0.01% + 0.005% 0.015% + 0.005% 0.015% + 0.005% 0.02% + 0.005%
1.1 V 10 µV 0.01% + 0.005% 0.015% + 0.005% 0.015% + 0.005% 0.02% + 0.005%
11 V 100 µV 0.01% + 0.005% 0.015% + 0.005% 0.015% + 0.005% 0.02% + 0.005%
Temperature Coefficient: (0.001% of output + 0.001% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Maximum Output Current: 10 mA
Loading: (0.001% f.s. + 1 µV)/ mA
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc